A template for linear calibration. Enter the spectrum id, the pixel distance to each peak (zero left), and the known wavelengths for the peaks. Wavelengths for two known peaks for CFL are entered.
The macro publishing and sharing system is under construction. To share your macros, please email the mailing list at plots-spectrometry@googlegroups.com -- ideally with a link to your macro on gist.github.com and it will be added.
Use a fluorescent bulb spectrum to calibrate your spectrometer
You can calibrate with a clear fluorescent spectrum like the one shown here. If the current spectrum is not usable, you can record a new one, and if you do not change your spectrometer, you can use the calibration for subsequent spectra.
First, click on the highest peak of the middle blue band:
Complete your calibration
Now, click at the peak of the green band as shown here:
Choose where to extract data from this image
To generate data, we look at the brightness of one slice of the image. Here, click on the image to choose which row to sample from.